Differences between revisions 3 and 18 (spanning 15 versions)
Revision 3 as of 2024-05-01 17:27:53
Size: 697
Comment:
Revision 18 as of 2024-05-21 17:00:44
Size: 3540
Comment:
Deletions are marked like this. Additions are marked like this.
Line 7: Line 7:
|| ''''' Description ''''' || ''''' links'''''|| ''''' Notes'''''||
|| RF system overview ||[[Electronics/RF_System]]|| ||
|| Generation box || [[Electronics/RF_System/frequency_generation_box]] || /!\ ''' Update me! '''||
|| ''''' Components ''''' || ''''' Quantity'''''|| ''''' Manufacturers '''''||
|| Pneumatic Valves || 7 || VAT ||
|| Manual Valves || 2 || VAT||
|| Turbo Molecular Pumps || 3 || Agilent (2) and OSAKA Vacuum (1) ||
|| Roughing Pumps || 6 || Agilent (1), Leybold (3), Varian (2)||
|| CRYOgenic Pumps || 1 || CTI-CRYOGENICS ||
|| Pirani Gauges || 12 || Super Bee (1), MKS(11)||
|| Full Range Gauges || 5 || Agilent ||
|| Gauge Controllers || 6 || Granville-Phillips (3), MKS(2). and Agilent (1)||
Line 18: Line 25:
|| ''''' Manufacturers ''''' || ''''' Model ''''' || ''''' Type ''''' || ''''' Pumping Speed ''''' || ''''' ''''' ||
|| Agilent || IDP 7 || Dry Scroll Pump || ||
|| Varian || IDP 7 || Dry Scroll Pump || ||
|| Varian || IDP 7 || Dry Scroll Pump || ||
|| Leybold || Manual Valves || Oiled Rotary Vane Pump || VAT||
|| || Turbo Molecular Pumps || 3 || Agilent (2) and OSAKA Vacuum (1) ||
Line 28: Line 40:
The 40m has an RGA SRS200. The volume for this RGA can be isolated using VM1, VM2, and VM3. This volume NEEDS TO REMAIN IN VACUUM IF THE RGA IS ON. If the filament is hot and exposed to Atmospheric pressure, the filament will burst and destroy the RGA. Keep th

|| ''''' Component ''''' || '''' Software '''' || ''''' Cylinders Ordered ''''' || ''''' Monthly Cost ''''' || ''''' Yearly Cost ''''' ||
|| RGA SRS 200 || || ||


= Nitrogen Cylinders =

The Nitrogen Cylinders are used to maintain the Open Status of the Pneumatic Valves. Without a pressure of 80 PSI in these lines, the pneumatic valves will close and the interlocks will trip. No valve can be opened until these interlocks have been switched off.

|| ''''' Component ''''' || '''' System Usage '''' || ''''' Cylinders Ordered ''''' || ''''' Monthly Cost ''''' || ''''' Yearly Cost ''''' ||
|| NI RP300 Research Grade Nitrogen || Top Gun || 1 || $252.58 || $252.58 ||

Recommended Items for the swap over to the Air Compressor

|| ''''' Component ''''' || ''''' Cost ''''' || ''''' Link ''''' ||
|| Super-Dry D-Series Desiccant Air Dryer || $1,109.85 || [[https://www.compressorworld.com/products/50379/super-dry-d-series-desiccant-air-dryer| AirDryer]] ||
|| Ultra Quiet and Oil-Free Air Compressor (Sound-Proof Case) || $1,150.00 || [[https://www.homedepot.com/p/California-Air-Tools-8-Gal-1-HP-Ultra-Quiet-and-Oil-Free-Air-Compressor-in-Sound-Proof-Cabinet-8010SPC/204768821| AirCompressor]] ||
Line 31: Line 61:
= Suggested Upgrades =
Line 32: Line 63:
= Nitrogen Cylinders = || ''''' Current Component '''''|| ''''' Model ''''' || ''''' Suggested Upgrade Options''''' || ''''' Link ''''' || ''''' Reasoning for upgrade ''''' ||
|| Leybold Rotary Vane Pump || D30A || || || The D30A pumps are Oiled Rotary Vane Pumps. After reading about the "Anti-Suck Back Mechanism" in the manual, it is shown that this "prevents pump fluid from surging back into the vacuum line" in Section 2.5 . This should be upgraded to Scroll pumps so that we do not run the risk of Oil back Streaming and destroying our entire Vacuum System.||
|| Pirani Gauges || 12 || Anest Iwata || || ||
|| Full Range Gauges || 5 || || || ||
|| Gauge Controllers || 6 || || || ||

This is the page for the 40m Vacuum System

Vacuum System as of 2024

Our full System contains the following components

Components

Quantity

Manufacturers

Pneumatic Valves

7

VAT

Manual Valves

2

VAT

Turbo Molecular Pumps

3

Agilent (2) and OSAKA Vacuum (1)

Roughing Pumps

6

Agilent (1), Leybold (3), Varian (2)

CRYOgenic Pumps

1

CTI-CRYOGENICS

Pirani Gauges

12

Super Bee (1), MKS(11)

Full Range Gauges

5

Agilent

Gauge Controllers

6

Granville-Phillips (3), MKS(2). and Agilent (1)

As of today (05/01/2024) the 40m Vacuum System does not use any of the CRYOgenic System. So the following components are apparent and physically attached to the system, but NOT in use.

Roughing Pumps

|| Manufacturers || Model || Type || Pumping Speed || ||

Agilent

IDP 7

Dry Scroll Pump

Varian

IDP 7

Dry Scroll Pump

Varian

IDP 7

Dry Scroll Pump

Leybold

Manual Valves

Oiled Rotary Vane Pump

VAT

Turbo Molecular Pumps

3

Agilent (2) and OSAKA Vacuum (1)

Turbo Molecular Pumps

Residual Gas Analyzer

The 40m has an RGA SRS200. The volume for this RGA can be isolated using VM1, VM2, and VM3. This volume NEEDS TO REMAIN IN VACUUM IF THE RGA IS ON. If the filament is hot and exposed to Atmospheric pressure, the filament will burst and destroy the RGA. Keep th

Component

' Software '

Cylinders Ordered

Monthly Cost

Yearly Cost

RGA SRS 200

Nitrogen Cylinders

The Nitrogen Cylinders are used to maintain the Open Status of the Pneumatic Valves. Without a pressure of 80 PSI in these lines, the pneumatic valves will close and the interlocks will trip. No valve can be opened until these interlocks have been switched off.

Component

' System Usage '

Cylinders Ordered

Monthly Cost

Yearly Cost

NI RP300 Research Grade Nitrogen

Top Gun

1

$252.58

$252.58

Recommended Items for the swap over to the Air Compressor

Component

Cost

Link

Super-Dry D-Series Desiccant Air Dryer

$1,109.85

AirDryer

Ultra Quiet and Oil-Free Air Compressor (Sound-Proof Case)

$1,150.00

AirCompressor

Suggested Upgrades

Current Component

Model

Suggested Upgrade Options

Link

Reasoning for upgrade

Leybold Rotary Vane Pump

D30A

The D30A pumps are Oiled Rotary Vane Pumps. After reading about the "Anti-Suck Back Mechanism" in the manual, it is shown that this "prevents pump fluid from surging back into the vacuum line" in Section 2.5 . This should be upgraded to Scroll pumps so that we do not run the risk of Oil back Streaming and destroying our entire Vacuum System.

Pirani Gauges

12

Anest Iwata

Full Range Gauges

5

Gauge Controllers

6

Vacuum (last edited 2024-05-29 17:38:13 by JancarlosanchezATligoDOTorg)