Complete power shutdown and start-up plan

There will be a power shutdown for 40m lab completely on Feb 5th (Sat), 2022 from 10am to 2pm. Following is the detailed plan for how to shutdown the whole lab and bring it back up online.

1. Shutdown

1.1. CDS Shutdown

1.1.1. Creating backups

1.1.2. Slow computers

1.1.2.1. Old computers

telnet into following computers:

shutdown

If the computers do not respond (in cases where you cannot telnet into them or maybe the above shutdown command does not work through telnet, it is untested), it is ok to hard reboot them. This can be done using the key on the crate. If no key, press the reset button or keep the ON/OFF button pressed until the machine powers down.

1.1.2.2. New computers

ssh into the following computers:

And do the following:

sudo systemctl stop modbusIOC.service
sudo shutdown

I suppose shutting down c1vac changes the state of the gate valves. But shutting down (power down) of the acromag targets will change the relay (or signal) states. It may close the valves.

1.1.3. Fast computers

ssh into following computers:

sudo /sbin/shutdown -h now

ssh into following computers:

And do the following:

rtcds stop --all
sudo /sbin/shutdown -h now

Note: c1lsc/c1sus/c1ioo are entangled and stopping the models on one of the machines will cause the others unresponsive. I'd recommend to run "shutdown -h now" at once on these three machines without stopping the RT processes. (KA)

1.1.4. Teststand computers

First ssh into c1teststand

ssh c1teststand

then ssh into the following computers in order and do the following:

sudo shutdown

And then shutdown c1teststand computer itself.

Switch off the chassis power buttons, AA/AI power buttons, and then switch off the power supply at the bottom of the stand.

1.1.5. Martian computers

1.1.5.1. Megatron

ssh megatron
sudo systemctl stop FSSSlow.service
sudo shutdown

1.1.5.2. Optimus

ssh optimus
cd /home/controls/Git/40m/scripts/MC
sudo docker-compose down
cd /home/controls/Git/40m/softchansmodbus
sudo docker-compose down
sudo shutdown

1.1.5.3. viviana

ssh viviana
sudo shutdown

1.1.5.4. paola

ssh paola
sudo shutdown

1.1.5.5. fb1

ssh fb1
sudo shutdown

1.1.5.6. chiara

ssh chiara
sudo shutdown

1.1.5.7. nodus

ssh nodus
sudo shutdown

1.2. Electronics racks

1.2.1. Yend Rack

If you switch off any of the power buttons on the rack, list them down here.

1.2.2. 1Y4

If you switch off any of the power buttons on the rack, list them down here.

1.2.2.1. Sorenson power supplies

Location

Model

Voltage

Ampere

1Y4-36

Sorensen DCS33-33E

5.0

0.8

1Y4-37

Sorensen DCS33-33E

5.0

0.8

1Y4-38

Sorensen DCS33-33E

15.0

1.6

1Y4-39

Sorensen DCS33-33E

15.0

1.5

1Y4-40

Sorensen DCS33-33E

20.0

0.3

1Y4-41

Sorensen DCS33-33E

20.0

0.3

1.2.3. 1Y3

If you switch off any of the power buttons on the rack, list them down here.

1.2.3.1. Sorenson power supplies

Location

Model

Voltage

Ampere

1Y3-39

Sorensen DCS33-33E

15.0

0.0

1Y3-40

Sorensen DCS33-33E

24.2

1.4

1Y3-GND

Agilent E3626A

15.0

0.0

1.2.4. 1Y2

If you switch off any of the power buttons on the rack, list them down here.

1.2.4.1. Sorenson power supplies

Location

Model

Voltage

Ampere

1Y2-39

Sorensen DCS33-33E

28.0

1.0

1Y2-40

Sorensen DCS33-33E

18.0

0.3

AUX-1Y2-0-1

Kepco High Voltage

5.0

0.5

AUX-1Y2-0-2

Kepco High Voltage

-5.0

0.0

AUX-1Y2-0-3

Kepco High Voltage

15.0

3.1

AUX-1Y2-0-4

Kepco High Voltage

15.0

3.6

AUX-1Y2-1-1

Kepco High Voltage

-15.0

2.4

AUX-1Y2-1-2

Kepco High Voltage

-15.0

1.9

AUX-1Y2-1-3

Kepco High Voltage

24.0

1.9

AUX-1Y2-1-4

Kepco High Voltage

-24.0

0.6

1.2.5. 1Y1

If you switch off any of the power buttons on the rack, list them down here.

1.2.6. 1Y0

If you switch off any of the power buttons on the rack, list them down here.

1.2.6.1. Sorenson power supplies

Location

Model

Voltage

Ampere

1Y0-1

Sorensen DCS33-33E

18.1

11.0

1Y0-2

Sorensen DCS33-33E

18.1

7.0

1.2.7. 1X1

If you switch off any of the power buttons on the rack, list them down here.

1.2.7.1. Power supplies

Location

Model

Voltage

Ampere

1X1-1

Sorensen DCS33-33E

5

1.6

1X1-2

Sorensen DCS33-33E

24

7.7

1X1-3

Sorensen DCS33-33E

24

2.9

1X1-4

Sorensen DCS33-33E

15

2.9

1X1-5

Sorensen DCS33-33E

15

2.3

1X1-6

Sorensen DCS33-33E

10

0.3

1X1-7

Sorensen DCS33-33E

5

2.2

1X1-9-1

Kepco High Voltage

160

0.025

1X1-9-1

Kepco High Voltage

160

0.045

1X1-9-1

Kepco High Voltage

160

0.020

1X1-9-1

Kepco High Voltage

100

0.010

1.2.8. 1X2

If you switch off any of the power buttons on the rack, list them down here.

1.2.9. AUX OMC SOUTH (Small rack south of 1X2)

If you switch off any of the power buttons on the rack, list them down here.

Location

Model

Voltage

Ampere

AUX-OMC-1

Sorensen DCS33-33E

18

0.3

AUX-OMC-2

Sorensen DCS33-33E

28

1.1

AUX-OMC-3

Sorensen DCS33-33E

18

0.6

AUX-OMC-4

Sorensen DCS33-33E

18

0.7

AUX-OMC-5

Sorensen DCS33-33E

12

0.6

AUX-OMC-6

Sorensen DCS33-33E

-

-

1.2.10. 1YNot

If you switch off any of the power buttons on the rack, list them down here.

1.2.11. 1X4

If you switch off any of the power buttons on the rack, list them down here.

1.2.12. 1X5

If you switch off any of the power buttons on the rack, list them down here.

1.2.12.1. Power supplies

Location

Model

Voltage

Ampere

1X5-36

Sorensen DCS33-33E

6

0

1X5-37

Sorensen DCS33-33E

5

0

1X5-38

Sorensen DCS33-33E

15

6.8 (found 6.6 while turning off)

1X5-39

Sorensen DCS33-33E

15

6.6

1X5-40

Sorensen DCS33-33E

19.9

5

1X5-41

Sorensen DCS33-33E

20

2.7

1.2.13. 1X6

If you switch off any of the power buttons on the rack, list them down here.

1.2.14. 1X7

If you switch off any of the power buttons on the rack, list them down here.

1.2.15. Teststand

1.2.16. 1X8 (Vacuum Rack)

If you switch off any of the power buttons on the rack, list them down here.

1.2.16.1. Power supplies

Location

Model

Voltage

Ampere

1X8-2

Sorensen DCS33-33E

24

1.0

1.2.17. 1X9

If you switch off any of the power buttons on the rack, list them down here.

1.2.17.1. Power supplies

Location

Model

Voltage

Ampere

1X9-34

Sorensen DCS33-33E

5

0

1X9-35

Sorensen DCS33-33E

5

0

1X9-36

Sorensen DCS33-33E

15

2.1

1X9-37

Sorensen DCS33-33E

15

2.0

1X9-38

Sorensen DCS33-33E

20

1.1

1X9-39

Sorensen DCS33-33E

19.9

0.3

1X9-40

Sorensen DCS33-33E

-

-

under

Kepco HV PS

100.15

0.005529

1.2.18. Xend Rack

If you switch off any of the power buttons on the rack, list them down here.

1.3. Lasers and other instruments

1.3.1. PSL

1.3.2. Vertex + PSL table

1.3.3. X end station

If you switch off any of the power buttons on the rack, list them down here.

1.3.4. Y end station

If you switch off any of the power buttons on the rack, list them down here.

1.3.5. Oplev Lasers

1.4. Control room equipment

2. Startup

2.1. Office space

2.2. Main network stations

2.3. Control room equipment

2.4. PSL + Vertex instruments

2.5. YEND and XEND instruments

2.6. Other instruments

2.7. Electronics racks

2.8. CDS Shutdown

2.8.1. Slow computers

2.8.2. Fast computers

Complete_power_shutdowwn_2022 (last edited 2022-02-07 18:46:14 by FranciscocarcobaATligoDOTorg)