Line 3: Line 3:
 * Turn OFF PZT high voltage
Line 4: Line 5:

= Alignment precedure =

 * Put input PZTs in center of range.

 * Place cage-mount aperture targets on BS and PRM cages

 * Using PZT mirrors ONLY, center beam in PRM aperture, and on PR2 folding mirror, and center PRM retro-reflection in YAW at PZT1/Faraday.

 * Adjust PRM pitch biases to center retro-reflection in pitch.

 * Add brass-mount apertures to ITMY and ETMY.

 * Unclamping and adjusting in yaw, adjust PR2 and PR3 folding mirrors such that beam is through BS, ITMY, and ETMY apertures.

 * Using PR2/3 pitch hysteresis, adjust pitch to center beam through BS/ITMY/ETMY apertures.

 * Adjust ITMY biases to center ITMY retro-reflection on BS.

 * Put apertures on ITMX and ETMX cages.

 * Using BS, center beam on ITMX/ETMX. NOTE: SHOULD NOT BE TOUCHING INPUT POINTING/PZTs.

 * Adjust ITMX biases to center ITMX retro-relfection on BS.

 * Should be able to see AS MICH flashes.

 * Adjust SR2, SR3 to center beams on SR3, SRM.

 * Adjust SRM for retro-reflection of AS beam at BS.
Line 7: Line 38:
 * Put input PZTs in center of range.
 * Make sure beams are centered on
 * Center beam on all AS optics

 * Make sure REFL is clear
   * dither PRM, see motion on AP tables
 * Make sure AS is clear
   * dither BS/ITM, see motion on AP tables

 * Using IPANG/POS pick-off mirrors, center beams on:
Line 11: Line 48:

 * Check all OpLevs centered, in and out of vacuum

 * Turn PZT high voltage OFF

Vent checklists

  • Turn OFF PZT high voltage

Alignment precedure

  • Put input PZTs in center of range.
  • Place cage-mount aperture targets on BS and PRM cages
  • Using PZT mirrors ONLY, center beam in PRM aperture, and on PR2 folding mirror, and center PRM retro-reflection in YAW at PZT1/Faraday.
  • Adjust PRM pitch biases to center retro-reflection in pitch.
  • Add brass-mount apertures to ITMY and ETMY.
  • Unclamping and adjusting in yaw, adjust PR2 and PR3 folding mirrors such that beam is through BS, ITMY, and ETMY apertures.
  • Using PR2/3 pitch hysteresis, adjust pitch to center beam through BS/ITMY/ETMY apertures.
  • Adjust ITMY biases to center ITMY retro-reflection on BS.
  • Put apertures on ITMX and ETMX cages.
  • Using BS, center beam on ITMX/ETMX. NOTE: SHOULD NOT BE TOUCHING INPUT POINTING/PZTs.
  • Adjust ITMX biases to center ITMX retro-relfection on BS.
  • Should be able to see AS MICH flashes.
  • Adjust SR2, SR3 to center beams on SR3, SRM.
  • Adjust SRM for retro-reflection of AS beam at BS.

Closeup checklist

  • Center beam on all AS optics
  • Make sure REFL is clear
    • dither PRM, see motion on AP tables
  • Make sure AS is clear
    • dither BS/ITM, see motion on AP tables
  • Using IPANG/POS pick-off mirrors, center beams on:
    • IPPOS
    • IPANG
  • Check all OpLevs centered, in and out of vacuum

  • Turn PZT high voltage OFF