| * 10-30-2018SV I learned it today that Serial Port to Ethernet Conversion is not reliable so scrap this MKS quote. I did not find any gauges with Ethernet output | * 10-30-2018SV I learned it today that Serial Port to Ethernet Conversion is not reliable so scrap this MKS quote. I did not find any gauges with Ethernet output. Now we have a clear direction. |
Fall 2018 Upgrade of 40m Vacuum System
Goals
- Install a network of Ethernet-capable vacuum gauges to replace the old serial-switchbox units.
- Replace the EPICS/hardware translation layer with Python-based code.
- Other hardware changes? Valves, pumps, computers?
good web site of gauges + : https://confluence.slac.stanford.edu/pages/viewpage.action?pageId=178520731#Gauges/Controllers-ColdCathode
List of all hardware that will interface with the new digital system
Component |
Notes |
New/Replacement/Carryover |
Link to product info |
|||
e.g., Ethernet gauge |
x-end |
replacement |
https://products.inficon.com/en-us/nav-products/product/detail/vgc50x/ |
|||
EthernetIP Gateway |
pumpspool |
option |
https://www.frederickscompany.com/wp-content/uploads/2018/06/mx_ag_eip_ds.pdf/ |
|||
Ethernet-Serialport Adapter |
2 purchased |
for gauge test Acromag |
||||
cold cathode gauges |
need 7 |
no serial port |
http://www.instrutechinc.com/cold-cathode-vacuum-gauge-hornet/ |
|||
cold cathode |
option |
gauge with RS485 |
||||
cold cathode 971 & 972 |
cc & or pirani |
with RS485 |
||||
- 10-30-2018SV I learned it today that Serial Port to Ethernet Conversion is not reliable so scrap this MKS quote. I did not find any gauges with Ethernet output. Now we have a clear direction.
10-31-2018 The primary purpose of this vent is to simply replace the VME system with the Acromag system and verify that we are at the same operating state as we are at now (so the pump and gauge hardware is NOT upgraded for now). This amounts to implementing Python logic for the INTERLOCKS and checking that they work. The zeroth level test can be done o the electronics bench, by just changing EPICS values for pressure indicators and seeing that the appropriate DAC/BIO channels are triggered. https://wiki-40m.ligo.caltech.edu/Vacuum_Interlocks?highlight=%28interlocks%29 |