Differences between revisions 33 and 38 (spanning 5 versions)
Revision 33 as of 2012-01-03 23:02:43
Size: 1066
Editor: localhost
Comment: converted to 1.6 markup
Revision 38 as of 2022-08-16 23:41:42
Size: 1329
Comment:
Deletions are marked like this. Additions are marked like this.
Line 5: Line 5:
= Locking =
Lock Settings for IMC.
Note C1:IOO-MC_FASTSW = 1 (Enable) and C1:IOO-MC_POL = 1 (Minus) is required state for autolocker to catch lock.
Line 6: Line 9:
{{attachment:IMC_Lock_Settings.png|IMC Lock settings}}
Line 7: Line 11:
= Mode Cleaner Model =
  . [[MC_model| MC model ]]
 

Mode Cleaner

  • The Input Mode Cleaner is a suspended triangular cavity. Each mirror of the triangle is 0.25 kg, 3" in diameter and 1" thick. The input mode cleaner length is ~13.5 m (27 m round trip), the Finesse is ~1460. Cavity pole is 3.79kHz.

Locking

Lock Settings for IMC. Note C1:IOO-MC_FASTSW = 1 (Enable) and C1:IOO-MC_POL = 1 (Minus) is required state for autolocker to catch lock.

IMC Lock settings

Mode Cleaner Model

Beam centering

Alignment Sensing and Control

Analog circuits

Documents and Links

  • LIGO-T020145-00-R, "Characterization and commissioning of digital suspensions for the 40m Mode Cleaner"

  • LIGO-T020147-00-R, "Characterizing the length sensing and control system of the mode cleaner in the LIGO 40m lab"

  • RF_ASC, "Interferometer Wave Front Sensing"

Mode_Cleaner (last edited 2022-08-16 23:41:42 by AnchalguptaATligoDOTorg)