Line 12: Line 12:
   * ''' Demodulation phase = 5.4 deg ''' (at which the I signal doesn't respond to the Michelson displacement).
   * ~ very tiny DC light is on the RFPD
   * ''' Demodulation phase = 17.6 deg ''' (at which the I signal doesn't respond to the Michelson displacement).
 <<BR>>
Line 16: Line 16:
   * clear    * ''' Demodulation phase = -38.050 deg '''
Line 20: Line 20:
 || AS55 || ~0 ||
 || REFL || 1.9 mW ||
 || AS55 || 0.02 mW ||
 || REFL || 644 mW ||

PRMI Noise budget

Measurements

  • 7/Mar/2012 by KI

Sensor in use

  • AS55_Q for MIHC
    • Gain of whitening filter = 30 dB for both the I and Q.

    • Demodulation phase = 17.6 deg (at which the I signal doesn't respond to the Michelson displacement).


  • REFL33_I for PRCL
    • Gain of whitening filter = 42 dB for both the I and Q.

    • Demodulation phase = -38.050 deg

DC light

  • port name

    DC power [mW]

    AS55

    0.02 mW

    REFL

    644 mW

Sidebands

  • The values noted here were estimated from the OSAs at the AS and REFL ports.

Optical gains

  • Parameters

    description

    values

    notes

    Estimated Optical gain

    80 uW/m

    Measured pk-pk Michelson error signal

    1000 counts

    Michelson was freely swinging

    Raw optical gain

    5.9052 x10^9 [counts/meter]

    = 4 * pi / lambda * (pk value of the error signal)

    Conversion loss of mixer

    6.5 dB

    from the spec sheet of JMS-1H

    ADC conversion

    1638.4 counts / V

    16 bit, +/-20 V

    Trans impedance at 55 MHz

    556.3 Ohm

    AS55 characterization

    Responsibity of diodes

    0.8 [A/W]

    from the spec sheet of CS3102

    Attenuation in the AS path

Servo filters

  • UGF = 55 Hz
  • Only FM5 (10:1000) was activated.
  • G = -2
  • No limiter in the digital path.
  • Feeding the signal back to BS.
  • Dewhitening filters in BS are ON.
  • Assumption : BS actuator has a pole at 1 Hz with Q = 5
  • Time delay = (without considering the effects from the analog circuits and the down sampleing filters.)

RAM effects

  • There is a big RAM signal in the I signal. The signals float from -100 counts to 0 counts as the Michelson freely swings.
    • The amount of the RAM looks changing as a function of time.
  • The Q signal looks having not so much RAM effect.

Interferometer_Characterization/PRMI_Noise_Budget (last edited 2012-03-08 23:13:09 by KiwamuizumiATligoDOTorg)