Size: 772
Comment:
|
Size: 947
Comment:
|
| Deletions are marked like this. | Additions are marked like this. |
| Line 1: | Line 1: |
| Effect of the end mirror curvature error to the contrast of ¥¥a Fabry-Perot based Michelson interferometer | = Effect of the end mirror curvature error to the contrast of ¥¥a Fabry-Perot based Michelson interferometer = |
| Line 3: | Line 3: |
| We found that one of the ETMs of the 40m interferometer could be delivered with a curvature error of 5¥% from the nominal radius. The effect of this curvature error to the contrast of a FPMI was investigated by analytical calculation and simulations with FINESSE. |
We found that one of the ETMs of the 40m interferometer could be delivered with a curvature error of 5¥% from the nominal radius. The effect of this curvature error to the contrast of a FPMI was investigated by analytical calculation and simulations with FINESSE. |
| Line 8: | Line 5: |
| Here is the document. | [attachment:curvature_mismatch.pdf Here] is the document. |
| Line 10: | Line 7: |
| For the parameter similar to the 40m interferometer with nominal end mirror curvature of 57 m, the contrast defect of the FPMI is given by: ¥begin{equation} dC = 3.61 ¥times 10^{-4} dR^{2}+O^{3}(dR)¥,¥,¥,, ¥end{equation} and this is dominated by the higher-order modes. The result was confirmed with simulations on FINESSE~¥cite{BIB-FREISE-FINESSE}. |
For the parameter similar to the 40m interferometer with nominal end mirror curvature of 57 m, the contrast defect of the FPMI is given by: . dC = 3.61 x 10^-4^ dR^2^ + O^3^(dR) and this is dominated by the higher-order modes. The result was confirmed with simulations on FINESSE. attachment:curvature_mismatch_plot.png Relationship between the curvature error and the contrast defect. Red dots show the simulated results. The blue line is the curve for the analytical calculation. |
Effect of the end mirror curvature error to the contrast of ¥¥a Fabry-Perot based Michelson interferometer
We found that one of the ETMs of the 40m interferometer could be delivered with a curvature error of 5¥% from the nominal radius. The effect of this curvature error to the contrast of a FPMI was investigated by analytical calculation and simulations with FINESSE.
[attachment:curvature_mismatch.pdf Here] is the document.
For the parameter similar to the 40m interferometer with nominal end mirror curvature of 57 m, the contrast defect of the FPMI is given by:
dC = 3.61 x 10-4 dR2 + O3(dR)
and this is dominated by the higher-order modes.
The result was confirmed with simulations on FINESSE.
attachment:curvature_mismatch_plot.png
Relationship between the curvature error and the contrast defect. Red dots show the simulated results. The blue line is the curve for the analytical calculation.