Line 2: Line 2:

The data for the following measurements is located at [https://nodus.ligo.caltech.edu:30889/svn/trunk/docs/upgrade08/RFsystem/RFphotodiodes/POX/ /svn/trunk/docs/upgrade08/RFsystem/RFphotodiodes/POX/]

POX

The data for the following measurements is located at [https://nodus.ligo.caltech.edu:30889/svn/trunk/docs/upgrade08/RFsystem/RFphotodiodes/POX/ /svn/trunk/docs/upgrade08/RFsystem/RFphotodiodes/POX/]

RF Transimpedance

attachment:transimpedance.png

Transimpedance at 11 MHz: 6.3 kOhms.

Noise

attachment:noise.png

The noise data was fit to attachment:fitfunc.png with the following results:

ZRF = (2.07 +- 0.01) kOhms

Id = (24.7 +- 0.6) mA

chi2 / ndf = 42

RF Behavior with No Light

attachment:timeseries.png

attachment:dark_spectrum.png

Picture of Photodiode

attachment:POX.jpg

Picture taken on 12/13/2010.