Differences between revisions 33 and 38 (spanning 5 versions)
Revision 33 as of 2009-12-17 01:45:40
Size: 1902
Comment:
Revision 38 as of 2010-01-19 00:05:23
Size: 2809
Editor: KiwamuIzumi
Comment:
Deletions are marked like this. Additions are marked like this.
Line 11: Line 11:
(The labels in this old schematic contains some mistakes. For instance: the ''RFPD Input'' label should be ''Servo Input'').
Line 17: Line 18:

''2010-01-18 '' - The phase shifter does not work. Also, Local Oscillator's signals at frequencies above 80MHz get distorted when passing through the phase-shifter's that is placed before the mixer. Either bypass it or fix it.

==== Servo Filter ====
This is the transfer function of the Servo Filter of the PDH Box as measured between the Servo IN and Servo OUT terminals of the box.

attachment: Old_PDH_box.png
Line 32: Line 40:
The UPDH boxes can be found inside the RF cabinet in the Y arm of the lab.
Line 33: Line 43:
== Servo Filters == == Servo Filters Comparison ==
Line 37: Line 47:

The way the filter's transfer function has been measured is by a swept sine between the "SERVO INPUT" and the "PIEZO DRIVE OUTPUT" connection on the box front panel. The spectrum analyzer used for the measurement is the SR785 and the source amplitude is set at 0.1V.

Universal PDH Box

Version I (2008)

Prototype


This is the schematic: [attachment:OldBoxSchematic.pdf LIGO D080285] (The labels in this old schematic contains some mistakes. For instance: the RFPD Input label should be Servo Input).

Evaluation:

2008-11-06 - Today I noticed a strange and unexpected effect on the box. I left it with the input signal plugged in and the output cable disconnected for a couple of days. that was not the first time it happened. Actually I've always done it after leaving the lab to avoid to have a railing signal acting on the laser for hours in case the PLL loses lock. This time though, when I tried to used the box after a couple of days, I found the the output to the piezo had a DC offset of about 3-5 V. Unplugging the input, and also the cable from the mixer to the filter connector connector, didn't cancel the offset and the only thing top do was to unplug the power from the box, wait for a few seconds and then reconnect it. After that, the offset disappeared.

2008-11-07 - What happened yesterday repeated today, even after just few hours of functioning. A RESET button would be nice.

2010-01-18 - The phase shifter does not work. Also, Local Oscillator's signals at frequencies above 80MHz get distorted when passing through the phase-shifter's that is placed before the mixer. Either bypass it or fix it.

Servo Filter

This is the transfer function of the Servo Filter of the PDH Box as measured between the Servo IN and Servo OUT terminals of the box.

attachment: Old_PDH_box.png

Version II (2009)


This is the schematic: [attachment:NewBoxSchematic.pdf LIGO D0901351]

This is how the new version looks like.

[attachment:DSC_1058.JPG Outside]

[attachment:DSC_1063.JPG Inside]

attachment:DSC_1058scaled.JPG

attachment:DSC_1063scaled.JPG

Evaluation (Rana, Dec 16 '09):

  • missing capacitor C25 and C29;
  • missing 50 Ohm terminator right after the mixer: it is currently before the Servo Input but it should be just after the mixer;
  • voltage dividers missing heat sinks;

The UPDH boxes can be found inside the RF cabinet in the Y arm of the lab.


Servo Filters Comparison

The New and the Old PDH box have different servo filters. Here they are compared:

attachment:NewandOlfFilterTF.png

The way the filter's transfer function has been measured is by a swept sine between the "SERVO INPUT" and the "PIEZO DRIVE OUTPUT" connection on the box front panel. The spectrum analyzer used for the measurement is the SR785 and the source amplitude is set at 0.1V.

Electronics/PDH_Universal_Box (last edited 2012-01-03 23:02:41 by localhost)