Differences between revisions 3 and 34 (spanning 31 versions)
Revision 3 as of 2008-09-25 22:17:15
Size: 220
Comment:
Revision 34 as of 2009-12-17 17:09:46
Size: 2172
Comment:
Deletions are marked like this. Additions are marked like this.
Line 1: Line 1:
## page was renamed from Electronics \ PDH Universal Box
## page was renamed from PDH Universal Box
Line 3: Line 5:
== PDH Universal Box Evaluation Page == == Universal PDH Box ==
Line 5: Line 7:
This pages is intended to list potential issues with the PDH NIM box prototype designed by Rich Abbott. === Version I (2008) ===
Prototype
----
This is the schematic: [attachment:OldBoxSchematic.pdf LIGO D080285]
Line 7: Line 12:
Evaluation:

''2008-11-06 '' - Today I noticed a strange and unexpected effect on the box. I left it with the input signal plugged in and the output cable disconnected for a couple of days. that was not the first time it happened. Actually I've always done it after leaving the lab to avoid to have a railing signal acting on the laser for hours in case the PLL loses lock. This time though, when I tried to used the box after a couple of days, I found the the output to the piezo had a DC offset of about 3-5 V. Unplugging the input, and also the cable from the mixer to the filter connector connector, didn't cancel the offset and the only thing top do was to unplug the power from the box, wait for a few seconds and then reconnect it. After that, the offset disappeared.

''2008-11-07 '' - What happened yesterday repeated today, even after just few hours of functioning. A RESET button would be nice.

=== Version II (2009) ===
Line 8: Line 20:
CategoryHomepage This is the schematic: [attachment:NewBoxSchematic.pdf LIGO D0901351]

This is how the new version looks like.

||<:> [attachment:DSC_1058.JPG Outside] ||<:> [attachment:DSC_1063.JPG Inside] ||
||<:> attachment:DSC_1058scaled.JPG ||<:> attachment:DSC_1063scaled.JPG ||

Evaluation (Rana, Dec 16 '09):
 * missing capacitor C25 and C29;
 * missing 50 Ohm terminator right after the mixer: it is currently before the Servo Input but it should be just after the mixer;
 * voltage dividers missing heat sinks;

-----
== Servo Filters ==
The New and the Old PDH box have different servo filters. Here they are compared:

attachment:NewandOlfFilterTF.png

The way the filter's transfer function has been measured is by a swept sine between the "SERVO INPUT" and the "PIEZO DRIVE OUTPUT" connection on the box front panel. The spectrum analyzer used for the measurement is the SR785 and the source amplitude is set at 0.1V.

Universal PDH Box

Version I (2008)

Prototype


This is the schematic: [attachment:OldBoxSchematic.pdf LIGO D080285]

Evaluation:

2008-11-06 - Today I noticed a strange and unexpected effect on the box. I left it with the input signal plugged in and the output cable disconnected for a couple of days. that was not the first time it happened. Actually I've always done it after leaving the lab to avoid to have a railing signal acting on the laser for hours in case the PLL loses lock. This time though, when I tried to used the box after a couple of days, I found the the output to the piezo had a DC offset of about 3-5 V. Unplugging the input, and also the cable from the mixer to the filter connector connector, didn't cancel the offset and the only thing top do was to unplug the power from the box, wait for a few seconds and then reconnect it. After that, the offset disappeared.

2008-11-07 - What happened yesterday repeated today, even after just few hours of functioning. A RESET button would be nice.

Version II (2009)


This is the schematic: [attachment:NewBoxSchematic.pdf LIGO D0901351]

This is how the new version looks like.

[attachment:DSC_1058.JPG Outside]

[attachment:DSC_1063.JPG Inside]

attachment:DSC_1058scaled.JPG

attachment:DSC_1063scaled.JPG

Evaluation (Rana, Dec 16 '09):

  • missing capacitor C25 and C29;
  • missing 50 Ohm terminator right after the mixer: it is currently before the Servo Input but it should be just after the mixer;
  • voltage dividers missing heat sinks;


Servo Filters

The New and the Old PDH box have different servo filters. Here they are compared:

attachment:NewandOlfFilterTF.png

The way the filter's transfer function has been measured is by a swept sine between the "SERVO INPUT" and the "PIEZO DRIVE OUTPUT" connection on the box front panel. The spectrum analyzer used for the measurement is the SR785 and the source amplitude is set at 0.1V.

Electronics/PDH_Universal_Box (last edited 2012-01-03 23:02:41 by localhost)