Line 5: Line 5:
 * [attachment:AgilentSAbasics.pdf Agilent - Spectrum Analyzer Basics]  * [attachment:AgilentSAbasics.pdf Agilent - Spectrum Analysis Basics]

Agilent Spectrum Analyzer AG4395a

Documents

  • [attachment:AG4395a.pdf AG4395a Manual]
  • [attachment:AgilentSAbasics.pdf Agilent - Spectrum Analysis Basics]

Noise Spectrum Data

  • The display of the SA can't show the square root symbol, but still in Noise format the units are to intended as [unit]/rtHz

  • Noise spectrum data as download with the GPIB interface. The downloaded data always refers to power spectrum - and not spectral density - independently from what the display is showing. to convert to spectral density, one has to divide by the bandwidth BW. (The BW can be found written inside the .par file produced by the netgpibdata.py downloading script).