Differences between revisions 7 and 9 (spanning 2 versions)
Revision 7 as of 2012-07-09 18:18:11
Size: 729
Comment:
Revision 9 as of 2017-11-10 20:33:05
Size: 757
Comment:
Deletions are marked like this. Additions are marked like this.
Line 6: Line 6:
Calibration Using Green = Calibration Using Green =
[[Calibration/SoCal]]

Calibration

I'm working on a simple model of the differential arm length sensing and control loop to guide work on real-time calibration. The model is based on Figure 6-1 in Rana's thesis. We will start tests for a real-time calibration system on one of the 40 m arm cavities using the digital control system. - Eric Hendries

= Calibration Using Green = Calibration/SoCal

OSEMs

The rough calibration of the OSEMs we have now (2009 - the Adhikari/Fritschel OSEMs from 1999) is 2V/mm.
Then the signal gets whitened in hardware with a 30,100:3 (poles:zeros) whitening filter with a gain of 1 at DC.
The ADC calibration is then 16384 cts/V.

Seismometers

Magnetometers

Steve

Calibration (last edited 2017-11-10 20:33:16 by CraigcahillaneATligoDOTorg)