This page should contain various calibration information.

Calibration Using Green

=== OSEMs ===
The rough calibration of the OSEMs we have now (2009 - the Adhikari/Fritschel OSEMs from 1999) is 2V/mm. [[BR]]
Then the signal gets whitened in hardware with a 30,100:3 (poles:zeros) whitening filter with a gain of 1 at DC. [[BR]]
The ADC calibration is then 16384 cts/V. [[BR]]

=== Seismometers ===

=== Magnetometers ===

=== Steve ===

Calibration

This page should contain various calibration information.

Calibration Using Green

OSEMs

The rough calibration of the OSEMs we have now (2009 - the Adhikari/Fritschel OSEMs from 1999) is 2V/mm. BR Then the signal gets whitened in hardware with a 30,100:3 (poles:zeros) whitening filter with a gain of 1 at DC. BR The ADC calibration is then 16384 cts/V. BR