Size: 556
Comment:
|
Size: 1996
Comment:
|
| Deletions are marked like this. | Additions are marked like this. |
| Line 1: | Line 1: |
| = X end laser = | <<TableOfContents(2)>> |
| Line 3: | Line 3: |
| || || Innolight Mephisto || || Part Number || || || Power || 1 W || || Nominal Current || 2.0 A || || Modulation Frequency fm|| 216.075 kHz || || PM/AM ratio at fm || 6e+5 [rad/RIN]|| |
= X end laser (Innolight) = == Nominal Parameters == || || Innolight Mephisto || || Part Number || || || Power || 1 W || || Nominal Current || 2.0 A || || Modulation Frequency fm|| 216.075 kHz || || PM/AM ratio at fm || 6e+5 [rad/RIN]|| |
| Line 12: | Line 14: |
| attachment:PMAM_innolight.png | . attachment:PMAM_innolight.png |
| Line 15: | Line 17: |
| == Temperature and beatnote == | == Temperature and beatnote (PSL vs X end laser ) == . attachment:temp_xend_psl.png == Laser Frequency vs Crystal Temperature == . Plots of the laser frequency vs temperature (measured by beating against a second laser) are shown below. [http://nodus.ligo.caltech.edu:8080/40m/2794 Also, see the elog here]. || '''Laser''' || '''df/dT''' || '''df/dV - within a mode''' || '''df/dV - average''' || || Innolight || 3.3MHz/mK || - || -|| . <<ImageLink(innolight_temperature_sweep.png,Upgrade 09/GreenLock/Control Information/innolight_temperature_sweep,height=400)>> |
| Line 20: | Line 35: |
| = Y end laser = || || LightWave || || Part Number || || || Power || 700m W (?)|| || Nominal Current || || || Modulation Frequency fm|| - || || PM/AM ratio at fm || - || |
-------- = Y end laser (Lightwave) = == Nominal Parameters == || || Lightwave || || Part Number || || || Power || 700m W (?)|| || Nominal Current || || || Modulation Frequency fm|| - || || PM/AM ratio at fm || - || |
| Line 30: | Line 48: |
| attachment:PMAM_lightwave.png | . attachment:PMAM_lightwave.png == Temperature and beatnote (PSL vs Y end laser ) == === Coarse scan === . attachment:temp_yend_psl_coarse.jpg === Fine scan === . attachment:temp_yend_psl.jpg === Fitting results === . Y_arm_Temp_meas = 0.95152*T_PSL + 3.8672 . Y_arm_Temp_set = 0.87326*T_PSL + 6.9825 == Laser Frequency vs Crystal Temperature == . Plots of the laser frequency vs temperature (measured by beating against a second laser) are shown below. [http://nodus.ligo.caltech.edu:8080/40m/2794 Also, see the elog here]. || '''Laser''' || '''df/dT''' || '''df/dV - within a mode''' || '''df/dV - average''' || || Lightwave || 2.8MHz/mK || 5GHz/V || 1.6GHz/V || . <<ImageLink(lightwave_temperature_sweep.png,Upgrade 09/GreenLock/Control Information/lightwave_temperature_sweep,height=400)>> [[BR]] |
Contents
X end laser (Innolight)
Nominal Parameters
Innolight Mephisto
Part Number
Power
1 W
Nominal Current
2.0 A
Modulation Frequency fm
216.075 kHz
PM/AM ratio at fm
6e+5 [rad/RIN]
Modulations
- attachment:PMAM_innolight.png
Temperature and beatnote (PSL vs X end laser )
- attachment:temp_xend_psl.png
Laser Frequency vs Crystal Temperature
Plots of the laser frequency vs temperature (measured by beating against a second laser) are shown below. [http://nodus.ligo.caltech.edu:8080/40m/2794 Also, see the elog here].
Laser
df/dT
df/dV - within a mode
df/dV - average
Innolight
3.3MHz/mK
-
-
Y end laser (Lightwave)
Nominal Parameters
Lightwave
Part Number
Power
700m W (?)
Nominal Current
Modulation Frequency fm
-
PM/AM ratio at fm
-
Modulations
- attachment:PMAM_lightwave.png
Temperature and beatnote (PSL vs Y end laser )
Coarse scan
- attachment:temp_yend_psl_coarse.jpg
Fine scan
- attachment:temp_yend_psl.jpg
Fitting results
- Y_arm_Temp_meas = 0.95152*T_PSL + 3.8672
- Y_arm_Temp_set = 0.87326*T_PSL + 6.9825
Laser Frequency vs Crystal Temperature
Plots of the laser frequency vs temperature (measured by beating against a second laser) are shown below. [http://nodus.ligo.caltech.edu:8080/40m/2794 Also, see the elog here].
Laser
df/dT
df/dV - within a mode
df/dV - average
Lightwave
2.8MHz/mK
5GHz/V
1.6GHz/V